Objective To investigate the computed tomography dose index (CTDI) of the single source fast kilovoltage-switching dual energy CT (SS FKS-DECT) ofthe abdomen compared to those of conventional single source single energy (SSSE) CT. ( Sakarya Med J 2018, 8(1):41-45 )Materials andMethodsIn this methodological research, the mean CTDI of the SS FKS-DECT of the abdomen obtained from 73 patients (32 women and 41 men)with a mean age of 5215.7 years, compared to the CTDI of the conventional SSSE CT. Body mass indices (BMI) of all patients and CTDI ofall scans were noted. Two-tailed student t test was used for statistical analysis. A P value of 0.05 was considered as statistically significant.Results The mean BMI was 24.27.6 and the mean CTDI for the FKS-DECT and SECT were 14.62.5 mGy and 12.57.2 mGy, respectively. Thedifference between the CTDI values of both acquisitions were statistically significant (p0.02). The mean CTDI of the FKS-DECT of theabdomen was14.4% higher than those for the SECT. .Conclusion The mean CTDI of the SS FKS-DE abdomen CT was 14.4% higher than those of the SS SE abdomen CT.
Amaç Tek kaynaklı tek enerji (TK TE) ve tek kaynaklı hızlı voltaj değişimli çift enerji (TK HVD-ÇE) abdomen bilgisayarlı tomografilerine (BT) ait BTdoz indekslerini (BTDİ) karşılaştırmak. ( Sakarya Tıp Dergisi 2018, 8(1):41-45 ).Gereç veYöntemlerBu metodolojik çalışmada, yaş ortalaması 5215,7 yıl olan 73 hastadan (32 Kadın ve 41 Erkek) elde olunmuş HVD-ÇE abdomen BT’lere aitortalama BTDİ, TE abdomen BT’lere ait ortalama BTDİ’ler ile karşılaştırıldı. Tüm hastaların vücut kitle indeksleri (VKİ) ve BT taramalarına aitBTDİ’leri kaydedildi. İstatistiksel analiz için örneklem t testi kullanıldı. P değeri 0.05 istatistiksel olarak anlamlı kabul edildi.Bulgular Ortalama VKİ 24.27.6 ve HVD-ÇEBT ile TEBT’lere ait ortalama BTDİ değerleri sırasıyla 14.62.5 mGy ve 12.57.2 mGy idi. Her iki teknikile elde olunan ortalama BTDİ’ler arasındaki fark anlamlıydı (p0.02). HVD-ÇE abdomen BT’ye ait ortalama BTDİ, TE abdomen BT’ye göre%14.4 daha fazla idi.Sonuç: HVD-ÇE abdomen BT’ye ait ortalama BTDİ, TE abdomen BT’ye göre %14.4 daha fazla idi.
Yazar |
Topcuoğlu, Osman Melih |
Yayın Türü | Article |
Tek Biçim Adres | https://hdl.handle.net/20.500.11831/2062 |
Konu Başlıkları | Genel ve Dahili Tıp |
Koleksiyonlar |
Araştırma Çıktıları | Ön Baskı | WoS | Scopus | TR-Dizin | PubMed 04- TR-Dizin İndeksli Yayınlar Koleksiyonu |
Dergi Adı | Sakarya Tıp Dergisi |
Cild | 8 |
Dergi Sayısı | 1 |
Sayfalar | 41 - 45 |
Yayın Tarihi | 2018 |
Eser Adı [dc.title] | Comparison Of Computed Tomography Dose İndices For Single Energy And Fast Kilovoltage Switching Dual Energy Abdominal Computed Tomographies |
Yazar [dc.contributor.author] | Topcuoğlu, Osman Melih |
Yayın Türü [dc.type] | article |
Özet [dc.description.abstract] | Objective To investigate the computed tomography dose index (CTDI) of the single source fast kilovoltage-switching dual energy CT (SS FKS-DECT) ofthe abdomen compared to those of conventional single source single energy (SSSE) CT. ( Sakarya Med J 2018, 8(1):41-45 )Materials andMethodsIn this methodological research, the mean CTDI of the SS FKS-DECT of the abdomen obtained from 73 patients (32 women and 41 men)with a mean age of 5215.7 years, compared to the CTDI of the conventional SSSE CT. Body mass indices (BMI) of all patients and CTDI ofall scans were noted. Two-tailed student t test was used for statistical analysis. A P value of 0.05 was considered as statistically significant.Results The mean BMI was 24.27.6 and the mean CTDI for the FKS-DECT and SECT were 14.62.5 mGy and 12.57.2 mGy, respectively. Thedifference between the CTDI values of both acquisitions were statistically significant (p0.02). The mean CTDI of the FKS-DECT of theabdomen was14.4% higher than those for the SECT. .Conclusion The mean CTDI of the SS FKS-DE abdomen CT was 14.4% higher than those of the SS SE abdomen CT. |
Özet [dc.description.abstract] | Amaç Tek kaynaklı tek enerji (TK TE) ve tek kaynaklı hızlı voltaj değişimli çift enerji (TK HVD-ÇE) abdomen bilgisayarlı tomografilerine (BT) ait BTdoz indekslerini (BTDİ) karşılaştırmak. ( Sakarya Tıp Dergisi 2018, 8(1):41-45 ).Gereç veYöntemlerBu metodolojik çalışmada, yaş ortalaması 5215,7 yıl olan 73 hastadan (32 Kadın ve 41 Erkek) elde olunmuş HVD-ÇE abdomen BT’lere aitortalama BTDİ, TE abdomen BT’lere ait ortalama BTDİ’ler ile karşılaştırıldı. Tüm hastaların vücut kitle indeksleri (VKİ) ve BT taramalarına aitBTDİ’leri kaydedildi. İstatistiksel analiz için örneklem t testi kullanıldı. P değeri 0.05 istatistiksel olarak anlamlı kabul edildi.Bulgular Ortalama VKİ 24.27.6 ve HVD-ÇEBT ile TEBT’lere ait ortalama BTDİ değerleri sırasıyla 14.62.5 mGy ve 12.57.2 mGy idi. Her iki teknikile elde olunan ortalama BTDİ’ler arasındaki fark anlamlıydı (p0.02). HVD-ÇE abdomen BT’ye ait ortalama BTDİ, TE abdomen BT’ye göre%14.4 daha fazla idi.Sonuç: HVD-ÇE abdomen BT’ye ait ortalama BTDİ, TE abdomen BT’ye göre %14.4 daha fazla idi. |
Kayıt Giriş Tarihi [dc.date.accessioned] | 2020-03-17 |
Yayın Tarihi [dc.date.issued] | 2018 |
Açık Erişim Tarihi [dc.date.available] | 2020-03-17 |
Dil [dc.language.iso] | eng |
Konu Başlıkları [dc.subject] | Genel ve Dahili Tıp |
Haklar [dc.rights] | info:eu-repo/semantics/openAccess |
Alternatif Başlık [dc.title.alternative] | Tek Enerji ve Hızlı Voltaj Değişimli Çift Enerji Abdomen Bilgisayarlı Tomografilerine Ait Bilgisayarlı Tomografi Doz İndekslerinin Karşılaştırılması |
ISSN [dc.identifier.issn] | 2146-409x |
Yayının ilk sayfa sayısı [dc.identifier.startpage] | 41 |
Yayının son sayfa sayısı [dc.identifier.endpage] | 45 |
Dergi Adı [dc.relation.journal] | Sakarya Tıp Dergisi |
Dergi Sayısı [dc.identifier.issue] | 1 |
Cild [dc.identifier.volume] | 8 |
Tek Biçim Adres [dc.identifier.uri] | http://www.trdizin.gov.tr/publication/paper/detail/TXpJeU1UTTFOUT09 |
Tek Biçim Adres [dc.identifier.uri] | https://hdl.handle.net/20.500.11831/2062 |